Humidity and Dew-Point Instruments
Please note: Michell Instruments adopts a continuous development program which sometimes
necessitates speci cation changes without notice. Please contact us for latest version.
www.michell.com
Precision Manufacturing
Dimensions
WM281
Digital Relative Humidity & Temperature Transmitter, Wall Mount
Accessories and Spare Parts
RS422/485 to PC (RS232) converter 330185
USB cable/software for con guration F035263
SS sintered lter H2
SS sintered lter with te on coating J2
Delrin cap slotted with SS mesh lter K7
Delrin cap slotted with PTFE lter, white Z7
Hygrosmart with Pt100 output (RH = 0-1 V) I7-0-00-1
You can check your hygrometer with the Control Kit HKC
which is based on the principle of non-saturated salt
solutions. Refer to technical data sheet CONTROL KIT
HKC
The WM281 relative humidity transmitter includes the
interchangeable Hygrosmart module. The interchangeable
module lets you recalibrate the transmitter simply by
replacing the sensor head with the Hygrosmart module. As a
result, maintenance costs are greatly reduced and down-time
is minimized.
Highlights
Analog and digital output standard
Based on the interchangeable Hygrosmart module
Analog output signals selectable through software
Metric or US measurement units selectable through
software
Technical Speci cations
Performance
Measurement range (RH) 0–100% RH
Measurement range (T) -30 to +70°C (-22 to +158°F)
Accuracy at 23°C (73°F)
Humidity
<±2% RH (5-95% RH)
Accuracy at 23°C (73°F)
Temperature
±0.4°C (±0.72°F)
Stability – RH Sensor <±1% RH/year
Response time – RH Sensor <10 sec typical (for 90% of the step
change)
Electrical output/input
Output signal con gurable on
request
0–1, 0–5, 0–10 V
0–20 mA, 4–20 mA, RS485
Supply voltage 15 - 27 V AC / 18 - 38 V DC
Load resistance Current output: R 500
Power consumption 1.7 W
Operating conditions
Operating temperature
Probe
Housing
Storage
-30 to +85°C (-22 to +185°F)
-30 to +70°C (-22 to +158°F)
-40 to +70°C (-40 to +158°F)
Mechanical speci cation
Ingress protection IP67
Material
Housing
Probe
Aluminum die casting
Delrin
Dimensions
Housing
Probe
120 x 120 x 49.5mm (4.72 x 4.72 x 1.94”)
L=100mm, ø 19mm (L=3.93”, ø0.74”)
Weight 450g (15.87oz)
Electrical connections Screw terminals